Micro Motion transmitter and HFO Flowmeter is a new type of mechatronics intelligent volumetric flowmeter, absorbed the advantages internal error correction through microcomputer system and realizing the high accuracy of flowmeter. Besides the error compensation of mechanical system, it can also increase the temperature compensation, pressure compensation and so on.

Micro Motion transmitter specifications:

One active 4–20 mA output: Not intrinsically safe

Isolated to ±50 VDC from all other outputs and earth ground

Maximum load limit: 820 ohms

Model 1500 and Model 1700 can report mass flow or volume flow

Model 2700 can report mass flow, volume flow, density, temperature, or drive gain

Output is linear with process from 3.8 to 20.5 mA, per NAMUR NE43 (February 2003)

1700 and 2700 flow transmitter condition: 

Self switching AC/DC input, automatically recognizes supply voltage

Complies with low voltage directive 2006/95/EC per EN 61010-1 (IEC 61010-1) with amendment 2; Installation (Overvoltage) Category II, Pollution Degree 2

AC power: 85 to 265 VAC, 50/60 Hz, 6 watts typical, 11 watts maximum

DC power:18 to 100 VDC, 6 watts typical, 11 watts maximum

Minimum 22 VDC with 1000 feet of 18 AWG (300 meters of 0.8 mm2) power-supply cable

At startup, transmitter power source must provide a minimum of 1.5 amperes of short-term current at a minimum of 18 volts at the transmitter's power input terminals

Fuse: IEC 127-1.25 non-serviceable fuse, slow blow